Fitting menu
Feedback Optimization
Select the Feedback Optimization screen to run the feedback test.
1. Click the Start feedback test button for the device/ear you would like
to assess.
2. Follow the on-screen instructions to perform feedback optimization.
The status of the test appears above the graph (complete, incomplete,
not run). In cases where the environment may have impacted the test
(e.g., excess noise present), the final results include a combination of
measured values and predicted feedback threshold values. The status
then indicates Incomplete. Repeating the test will replace any predicted
values with measured values if they can be reliably obtained in the
current test environment.
Once the feedback test is completed, the hearing instrument’s gain is
limited as shown on the graph where:
• Black line = feedback threshold
• Grey line = hearing instrument gain limit
• Green line = target gain for 50 dB pure tone input
• Red or Blue line = aided gain for 50 dB pure tone input
Note: If desired, the feedback test can be deleted using the Delete
icon above the graph for the device / ear in question.
Note: The phase canceller is enabled prior to running the feedback
optimization test and the quick fit will be limited to the predicted
maximum stable gain. Following the feedback test, the device is
automatically quick fitted up to the desired adaptation manager level
setting, but not beyond the measured maximum stable gain curve.